Which freezing equipment does Scala Scientific supply?
Scala Scientific supplies freezing equipment for laboratories, hospitals and companies that need to store materials for a long time. Our range consists of seven product groups:
Which storage temperature do you need?
The temperature determines how long material stays usable. In a refrigerator at 5 °C it usually keeps for a few days. In a -20 °C freezer it often lasts several months up to a year, and ultra-low temperature freezers are intended for storage over a very long period. The difference between the inside of an ultra-low freezer and the lab is about 100 °C, which places high demands on insulation and on discipline in use.
Automatic or manual defrosting
Freezers in the range of -20 to -30 °C often have automatic defrosting: four times a day the evaporator is briefly heated so that ice melts. You do not notice this in the lab. Low and ultra-low temperature freezers (-40, -60, -80 and -85 °C) do not have this. When the door is opened, warm, humid air enters and settles as snow and ice. With good use of the snow brush and ice scraper that come with a new freezer, a freezer can go on for a few years. After that, defrosting is needed, usually after 12 to 18 months depending on use. If you do not have an empty freezer available to store the material in, you can rent one from us.
Placing a freezer in the lab
A freezer gives off heat, so the lab needs ventilation or air conditioning. Preferably do not place the freezer in direct sunlight and keep 10 to 15 cm of space free behind it. For -80 °C freezers, manufacturers also advise about 10 cm at the sides. Freezers with low energy consumption usually give off less heat as well. Several freezers side by side do not give off more heat in total than separately; -20 °C freezers can often be placed tightly next to each other.
Qualification, maintenance and service
For laboratories in pharma, biotech and food technology we carry out installation qualification and operational qualification (IQ/OQ) on request, based on your user requirement specification (URS). This is not a formality: the outcome is pass or fail, and any deviation leads to adjustment and requalification. For maintenance, preventive maintenance once every 12 months is a good rhythm for freezers in 9 out of 10 cases.
We aim to deliver within two weeks. With a maintenance contract (usually preventive maintenance once every 12 months) you can reach us day and night, without higher service rates outside office hours, and you can rent a replacement unit. The warranty is at least two years.
In the event of a failure, samples can warm up quickly. That is why we can supply a replacement freezer that is already at temperature in our warehouse.
Help with choosing
We work with 24 manufacturers and have advised laboratories for almost 18 years. Contact us for advice and a quotation. Also see our cooling equipment for storage above freezing point.